Korisnici mogu sami provjeriti kvalitet pH elektrode prema podacima u tabeli ispod, koraci su sljedeći (podešavanje pH metra na mV opseg)


⑴ Test the zero potential pH value, zero point pH value: 7±0.5pH pH electrode inserted into the pH4.00 buffer solution, read after stabilization (pre-test the temperature of the solution). For example, at 25°C, the reading value E1'=186mV (value), and the accurate value is 178mV (see table), then the error value=186-178mV, converted into pH value: 8/59.16=0.14pH (59.16mV/pH is K value at 25°C, see table). Therefore the zero pH value is acceptable. Generally, the zero potential error of a new electrode is ≤±0.3pH, if the zero potential error is >±0.5pH, greška ispitivanja elektrode će biti relativno velika, posebno kada se temperatura ispitne otopine jako promijeni. Ako tačnost testa nije visoka, vrijednost nulte potencijalne greške može biti veća, ali maksimum ne može preći 1pH.


⑵ postotak teorijske stope viška testne elektrode (PTS)


Zabilježite podatke testa u koraku (1) od 186mV, očistite elektrodu i ubacite je u otopinu pufera pH9,18 i očitajte nakon stabilizacije, E2'=111mV (vrijednost), ΔE'= E1' plus E2'=186 plus 111= 297mV, u poređenju sa ΔE u tabeli, PTS=97 posto, kvalifikovano, općenito je PTS novih elektroda veći ili jednak 97 posto , a elektrode koje se koriste duže vrijeme mogu biti opuštene, ali općenito PTS veći od ili jednak 95 posto, osim ako su zahtjevi za preciznost ispitivanja vrlo niski.


(3) Očistite elektrodu i zatim ubacite pH4.00 pufer otopinu, uporedite očitanje sa podacima u koraku (1), greška bi trebala biti manja ili jednaka ±2mV, u suprotnom to znači da je ponovljivost elektroda je loša.


⑷Probno očitavanje elektrode treba biti stabilno u roku od 30-60 sekundi, inače to znači da je odgovor elektrode presporo

 