The multimeter has an external trigger function. When the Ext Trig BNC terminal on the rear panel of the instrument receives a low-true TTL pulse to trigger the instrument (pulse width greater than 1 us), the multimeter can cache an external trigger. That is, if the multimeter is reading a reading and an external trigger A trigger occurs, the trigger is accepted and saved, and the trigger is accepted after the multimeter has finished taking the reading. To select the external trigger source through front panel operation, use the "Single" button on the front panel. To select the external trigger source by remote operation, use the TRIGger:SOURce EXTernal command. To set the number of readings to be read back in one trigger, use the SAMPle:COUNT command.