Kako pravilno skladištiti i koristiti pH pufer otopine?
After preparing the buffer solution, it should be placed in a glass bottle or a polyethylene bottle (alkaline pH buffer solutions such as pH 9.18, pH 10.01, pH 12.46, etc. should be placed in polyethylene bottles). The bottle cap should be tightly closed and stored at low temperature (5-10 ℃) in the refrigerator. It can generally be used for about six months. If turbidity, mold or precipitation are found, it cannot be continued to be used. When using, prepare several 50mL polyethylene small bottles, pour the buffer solution from the large bottles into the small bottles, and let them stand at ambient temperature for 1-2 hours until the temperature is balanced before use. After use, it should not be poured into the large bottle to avoid contamination. The buffer solution in the bottle can be used for 2-3 days under environmental conditions of>10 stepeni. Općenito, otopine pH 7,00, pH 6,86 i pH 14,00 mogu se koristiti duže vremenske periode. Otopine pH 9,18 i pH 10,01 sklone su pH promjenama zbog apsorpcije ugljičnog dioksida iz zraka.
Koja je svrha pH puferske otopine?
Kalibrirajte pH metar prije mjerenja pH.
Da biste provjerili tačnost pH metra, na primjer, nakon kalibracije pH metra s pH 6,86 i pH 14,00, umetnite pH elektrodu u otopinu pH 9,18 i provjerite da li je prikazana vrijednost instrumenta u skladu s pH vrijednošću standardne otopine.
Da li je potrebno ponovo kalibrirati pH metar tokom opšteg preciznog merenja. Nakon kalibracije i korištenja pH metra može doći do pomaka ili promjena. Stoga, prije testiranja, umetnite elektrodu u standardnu pufersku otopinu koja je blizu izmjerene otopine i odredite da li je potrebna ponovna kalibracija na osnovu veličine greške.